This is a guide for the pre-installation requirements of Thermo Scientific TRACE 1300 and TRACE 1310 gas chromatographs
This document serves as a Preinstallation Requirements Guide for the Thermo Fisher Scientific TRACE 1300 and TRACE 1310 Gas Chromatographs. It details the necessary site preparations, including laboratory remodeling, power supply specifications, environmental controls, and gas line requirements, to ensure a successful installation. The guide emphasizes that the site must be properly prepared to avoid additional costs for field service engineers.
| System Mass | 107 kg (236 lbs) |
|---|---|
| Operating Temperature | 15 and 27 °C (59 and 81 °F) |
| Relative Humidity | 40% and 80% |
| Doorway Width | 80 cm (32 in.) |
The minimum width required is 80 cm (32 in.).
The workbench must be sufficient for all equipment, support the system mass of 107 kg (236 lbs), and be free from vibration.