The document is the instruction manual of Signet 2551 magnetic flow meter
This manual provides instructions for installing, wiring, calibrating, and maintaining the Signet 2551 magnetic flow meter. It includes safety guidelines, hardware configuration, electrical specifications, and compatibility details. The meter is designed for insertion into pipes with diameters from DN15 to DN900 and supports frequency or 4-20 mA output. It features corrosion-resistant materials and is certified to CE, UL/CUL, and RoHS standards.
The manual also covers safety precautions, such as depressurizing the system before installation, using appropriate fittings, and grounding requirements. It includes technical specifications for electrical performance, environmental conditions, and material compatibility. Optional digital (S3L) output is available for integration with Signet controllers. This manual is written in Italian.
| Flow Range | DN15 to DN900 (0.5 to 36 in.) |
|---|---|
| Minimum Flow Velocity | 0.05 m/s (0.15 ft/s) |
| Linearity | ±(1% della lettura + 0,01 m/s) |
| Repeatability | ±0.5% della lettura a 25 °C |
| Minimum Conductivity | 20 μS/cm |
| Max Operating Pressure | 10.3 bar at 25 °C (150 psi at 77 °F) |
| Enclosure Rating | NEMA 4X/IP65 |
| Output Options | Frequency or 4-20 mA |
The materials include PP/316 SS, PVDF/Hastelloy-C, and PVDF/Titanium, depending on the model variant.
The maximum operating pressure is 10.3 bar at 25 °C (150 psi at 77 °F).
The Signet 2551 is available with frequency or 4-20 mA output. Digital (S3L) output is included for models with frequency output.
The minimum conductivity is 20 μS/cm.