This document is about Signet 2610 process optical dissolved oxygen sensor communication protocol manual, which introduces how to connect the sensor to a controller or recorder via RS485 Modbus.
This document serves as a communication protocol manual for the Signet 2610 Process Optical Dissolved Oxygen Sensor. It details how the sensor connects to controllers or recorders via RS485 Modbus, including power connection instructions and default settings for analog outputs.
The manual provides comprehensive tables for communication protocol registers and calibration registers. It defines specific register addresses for reading and writing data such as dissolved oxygen concentration, temperature, salinity, and atmospheric pressure. Additionally, it outlines a step-by-step procedure for calibrating the sensor using 100% and 0% saturation environments. This manual is written in Chinese.
| Connection Interface | RS485 Modbus |
|---|---|
| Default Device Address | 1 |
| Default Communication Baud Rate | 19200 |
| Data Quality ID Default | 0x0030 (48) |
| Default Atmospheric Pressure | 1013.25 mbar |
The default unit for dissolved oxygen concentration is mg/L.
Real-time salinity values must be written in PSU units ranging from 0 to 42 PSU.
Real-time atmospheric pressure values must be written in millibar, ranging from 506.625 to 1114.675 mbar.
No, regular calibration is not required as the sensor cap is factory calibrated to 2% accuracy.