Georg Fischer Signet 2450 Pressure Sensor User Manual

The Signet 2450 is a high-precision pressure sensor with a full PVDF construction for corrosive liquid applications. The sensor provides 4-20 mA and S3L digital outputs and has three pressure ranges for optimal measurement resolution. The sensor also features temperature compensation for exceptional accuracy over a wide operating range.


Georg Fischer Signet 2450 Pressure Sensor User Manual - cover page
Brand
Georg Fischer
Document type
User Manual
Language
Spanish
Pages
8
File format
PDF
File size
380 KB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
C70E168477D0FBF516C848EC9B508E14

About this manual

This manual provides installation instructions and technical specifications for the Signet 2450 pressure sensor, including compatibility with 4-20 mA or digital S3L outputs. It covers mounting options, material compatibility, and safety warnings. The sensor features a PVDF body and ceramic diaphragm for corrosion resistance and includes temperature compensation for accuracy. Installation requires proper venting, avoiding over-tightening, and using appropriate fittings to prevent damage.

Safety precautions include using personal protective equipment, avoiding static discharge, and ensuring no air bubbles in the tubing to prevent measurement errors. The sensor is rated for -15°C to 85°C operation and complies with CE, RoHS, ISO 9001, ISO 14001, and OHSAS 1800 standards. This manual is written in Spanish.

What's inside

  • Installation
  • Specifications
  • Safety Precautions
  • Mounting Options
  • Material Compatibility
  • Temperature Compensation
  • Wiring and Connections
  • Warranty Information
  • Product Registration

Specifications

Accuracy±1% of full scale at 25°C
Temperature Coefficient±0.03% of full scale per °C
Pressure Ranges-XU: 0-0.7 bar, -XL: 0-3.4 bar, -XH: 0-17.2 bar
Operating Temperature-15°C to 85°C
Casing MaterialPVDF
Diaphragm MaterialCeramic
Cable Type3 conductors + shield, 22 AWG, black/red/white/shield
Weight0.3 kg (0.65 lb)
Electrical SupplyDigital (S3L): 5V DC ±10%, 4-20mA: 12-24V DC ±10%

Frequently asked questions

What are the pressure ranges available for the Signet 2450?

The sensor is available in three pressure ranges: -XU (0-0.7 bar), -XL (0-3.4 bar), and -XH (0-17.2 bar).

What is the operating temperature range?

The sensor operates from -15°C to 85°C.