This manual provides installation, electrical connection, menu configuration, calibration procedures, and troubleshooting for the GEORG FISCHER SIGNET…
The document serves as an operation manual for the +GF+SIGNET 8850-1 Conductivity/Resistivity Transmitter. It covers two installation types: panel-mounted units which include necessary accessories, and field-installed units requiring a separate mounting kit (3-8050). Detailed instructions are provided for panel mounting using a 1/4 DIN hole saw and specific clearance distances.
The manual details the product specifications, including electrical performance such as power requirements (12~24VDC), sensor input ranges for conductivity and resistivity, and output signals. It explains how to connect the device via terminals for system power, sensor inputs, and open collector outputs. Configuration is managed through a menu system accessible with a specific key code, allowing users to set parameters like cell constant, units, temperature compensation, and calibration points. This manual is written in Chinese.
| Power Supply | 12~24VDC±5% |
|---|---|
| Conductivity Range | 0.055~400,000uS |
| Resistivity Range | 10KΩ ~18.2MΩ |
| Temperature Range | -25~120℃ |
| Operating Temperature | -10~70 ℃ |
| Display Type | LCD 2x16 bit |
A net margin of 1in. (25mm) should be reserved on each side of the instrument.
The maximum pulse rate is 400 pulses per minute.
The password consists of pressing the Up key three times followed by the Down key once.