Georg Fischer 2450 Pressure Sensor Manuals

ManualLib has 2 documents for the Georg Fischer 2450 Pressure Sensor: 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
2450 Pressure Sensor
Documents
2
Languages
EN · ES
Last added
2023-10-20

Start here

Georg Fischer Signet 2450 Pressure Sensor User Manual

Georg Fischer Signet 2450 Pressure Sensor User Manual

User Manual 8 pages · 414 KB · EN

This manual provides installation and wiring instructions for the Signet 2450 Pressure Sensor, including union and in-line mounting options. It details electrical specifications, safety precautions, and compatibility with GF transmitters. The sensor features a PVDF body and ceramic diaphragm for corrosion resistance and is available with digital or 4-20mA outputs.

Safety and installation guidelines emphasize depressurizing systems, avoiding overpressure, and using proper PPE. The manual also covers environmental limits, proof pressure ratings, and wiring configurations for both digital and analog outputs.

Read online

Georg Fischer 2450 Pressure Sensor specifications

Sensor Accuracy±1% of full scale @ 25 °C
Thermal Sensitivity Shift0.03% of full scale per °C
Max Loop Impedance (4-20mA)1000Ω @12V, 3250Ω @18V, 600Ω @24V
Power Requirements (Digital)5 to 6.5 VDC +10%, <1.5 mA
Power Requirements (4-20mA)12 to 24 VDC +10% regulated
Cable Length4.6 m (15 ft)
Wetted MaterialsSensor housing: PVDF; Diaphragm: Ceramic; Seal: FKM
Process Connection1/2 in. Union male thread or 3/4 in. NPT male thread

Source: Georg Fischer Signet 2450 Pressure Sensor User Manual

All Georg Fischer 2450 Pressure Sensor documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What are the pressure ranges for the 2450 Pressure Sensor?

The sensor offers three pressure ranges: -0.1 to 0.7 bar (–1.5 to 10 psi) for -XU, -0.4 to 3.4 bar (–6 to 50 psi) for -XL, and -0.96 to 17.2 bar (–14.6 to 250 psi) for -XH.

From: Georg Fischer Signet 2450 Pressure Sensor User Manual

What is the maximum proof pressure for the 2450 sensor?

The proof pressure is 150 bar (2175 psi) for -XU, 5.2 bar (75 psig) for -XL, and 20.7 bar (300 psig) for -XH.

From: Georg Fischer Signet 2450 Pressure Sensor User Manual

What are the pressure ranges available for the Signet 2450?

The sensor is available in three pressure ranges: -XU (0-0.7 bar), -XL (0-3.4 bar), and -XH (0-17.2 bar).

From: Georg Fischer Signet 2450 Pressure Sensor User Manual

What is the operating temperature range?

The sensor operates from -15°C to 85°C.

From: Georg Fischer Signet 2450 Pressure Sensor User Manual