ROSS sodium ion selective electrode user manual
This document serves as a comprehensive user guide for the ROSS Sodium Ion Selective Electrode, covering its preparation, operation, and maintenance. It details the necessary equipment and solutions required for accurate sodium ion measurement in aqueous solutions. The guide emphasizes the use of the known addition procedure for consistent and fast results.
| Response Time | Faster response than conventional sodium electrodes |
|---|---|
| Slope Range | 54 to 60 mV |
| Storage Solution | 5 M NaCl, 0.08 M NH4OH, and 0.08 M NH4Cl |
For the most consistent and fastest results, we recommend measuring sodium using the known addition procedure.
The electrode should be soaked for at least two hours in the sodium electrode storage solution.
The slope difference between two potential readings should be in the range of 54 to 60 mV at 20-25 ºC.