REED Instruments R6015 Wood Moisture Meter Instruction Manual

This manual describes the features, specifications, operating instructions, temperature correction, automatic power-off setup, calibration check, care and maintenance, calibration table for wood and common timber species of the R6015 wood moisture meter.


REED Instruments R6015 Wood Moisture Meter Instruction Manual - cover page
Brand
Reed Instruments
Category
Other
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
1.3 MB
Published
19 October, 2023
Updated
19 October, 2023
Model
R6015
MD5 checksum
F43E517C1EC3FCDE8B98893790A11262

About this manual

The REED Instruments R6015 Wood Moisture Meter is a conductivity-based device designed for the timber industry to measure moisture content in wood. It features eight calibration scales covering approximately 170 species of wood and includes automatic temperature compensation when used with the included temperature probe. The instrument allows for measurements using either built-in needle electrodes or a remote hand-press probe for hard surfaces.

What's inside

  • Specifications
  • Operating Instructions
  • Taking a Measurement
  • Measuring Without Temperature Probe
  • Automatically Temperature Corrected (ATC)
  • Auto Power-Off Set-up
  • Calibration Check
  • Care and Maintenance
  • Calibration Table for Wood
  • Common Timber Species

Specifications

Measuring PrincipleElectrical resistance
Measuring Range6 to 99.9% MC
Accuracy±1% MC
Moisture Electrode8mm long replaceable contact pins
Temp. Compensation Range-31 to 176°F (-35 to 80°C)
Power Supply2 x AAA batteries
Dimensions7x 1.9 x 2” (180 x 50 x 31mm)
Weight6.1 oz (175g)

Frequently asked questions

How many wood species can the R6015 measure?

The instrument has eight calibration scales enabling measurements in approximately 170 species of wood.

How do I switch the instrument on and off?

The instrument is switched ON by pressing the (!) button and switched OFF by pressing the (!) button for approx 3 seconds.

What is the default auto power-off time?

The default auto power-off time is 5 minutes, but it can be set manually from 1 to 9 minutes or disabled.

What is the acceptable range for a calibration check?

At the 'A' calibration, the instrument should register %H,O values in the range 17.7 to 18.3.