About this manual

This manual provides step-by-step instructions for operating the R6015 wood moisture meter, including how to take measurements with or without a temperature probe, auto power-off settings, and calibration checks. It also includes a calibration table for 150 wood species and care instructions for maintaining the device.

The meter uses electrical resistance to measure moisture content from 6% to 99.9% MC with ±1% accuracy. It features automatic temperature compensation, 8 wood calibration scales, and a built-in 2-point calibration check.

What's inside

  • Taking a Measurement
  • Measuring Without Temperature Probe
  • Automatically Temperature Corrected (ATC)
  • Auto Power-Off Set-up
  • Calibration Check
  • Care and Maintenance
  • Calibration Table for Wood
  • Common Timber Species

Specifications

Measuring PrincipleElectrical resistance
Measuring Range6 to 99.9% MC
Accuracy±1% MC
Moisture Electrode8mm long replaceable contact pins
Temp. Compensation-31 to 176°F (-35 to 80°C)
Power Supply2 x AAA batteries
Dimensions7x 1.9 x 2” (180 x 50 x 31mm)
Weight6.1 oz (175g)

Frequently asked questions

How do I calibrate the R6015 moisture meter?

Use the two calibrators in the cap, select scale A, disconnect the temperature probe, and touch the needle to the poles. If calibrated correctly, the meter should register %H,O values between 17.7–18.3 or 25.5–26.5.

What is the measuring range of the R6015?

The meter measures moisture content from 6% to 99.9% MC with ±1% accuracy.

Can the R6015 measure wood without a temperature probe?

Yes, but measurements need manual correction: subtract 0.5% MC per 5°C above 20°C, or add 0.5% MC per 5°C below 20°C.