Key facts
- Brand
- Reed Instruments
- Model
- R6015
- Documents
- 4
- Languages
- EN
- Last added
- 2023-10-19
Start here

REED Instruments R6015 Wood Moisture Meter Instruction Manual
This is the instruction manual for the REED Instruments R6015 wood moisture meter. It explains how to operate the device, including measuring moisture content using built-in pins or a remote probe, automatic temperature compensation, calibration checks, and auto power-off settings. The manual also includes a calibration table for common timber species and maintenance tips.
The device uses electrical resistance to measure moisture content in wood from 6% to 99.9% MC with ±1% accuracy. It includes a temperature probe for automatic compensation and features 8 calibration scales for 170 wood species. The meter runs on 2 x AAA batteries and comes with a carrying case and accessories.
REED Instruments R6015 specifications
| Measuring Principle | Electrical resistance |
|---|---|
| Measuring Range | 6 to 99.9% MC |
| Accuracy | ±1% MC |
| Moisture Electrode | 8mm long replaceable contact pins |
| Temp. Compensation | -31 to 176°F (-35 to 80°C) |
| Power Supply | 2 x AAA batteries |
| Dimensions | 7 x 1.9 x 2” (180 x 50 x 31mm) |
| Weight | 6.1 oz (175g) |
Source: REED Instruments R6015 Wood Moisture Meter Instruction Manual
All REED Instruments R6015 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How do I measure moisture without a temperature probe?
Measure at 20°C (68°F) and manually adjust by ±0.5% per 5°C deviation from 20°C.
From: REED Instruments R6015 Wood Moisture Meter Instruction Manual
How do I set the auto power-off time?
Press and hold the button and press the button to set auto power-off from 1 to 9 minutes or disable it (0).
From: REED Instruments R6015 Wood Moisture Meter Instruction Manual
How do I calibrate the R6015 moisture meter?
Use the two built-in calibrators in the cap. Select scale A, disconnect the temperature probe, and touch the needle to the calibrator poles. The meter should register 17.7–18.3% H2O at 'T' or 25.5–26.5% at 'B'.
What is the auto power-off setting?
The default auto power-off is 5 minutes, but it can be set from 1 to 9 minutes or disabled by pressing and holding the button and pressing the button to change the code.
How do I calibrate the R6015 moisture meter?
Use the two calibrators in the cap, select scale A, disconnect the temperature probe, and touch the needle to the poles. If calibrated correctly, the meter should register %H,O values between 17.7–18.3 or 25.5–26.5.
From: REED Instruments R6015 Wood Moisture Meter Instruction Manual
What is the measuring range of the R6015?
The meter measures moisture content from 6% to 99.9% MC with ±1% accuracy.
From: REED Instruments R6015 Wood Moisture Meter Instruction Manual