The Newport Vibration Control System provides an ideal working platform for vibration-sensitive devices such as interferometers, microscopes, and balances. The system integrates Newport's rigid, laminated honeycomb panel technology and pneumatic isolation systems to provide a mounting platform which is rigid, yet thin and light weight.

The document covers the complete lifecycle of the system, including unpacking and inspection, assembly of the table and isolators, leveling procedures, and operation principles. It also includes troubleshooting guides for common issues like poor isolation performance or air leaks, as well as maintenance and factory service information.

What's inside

  • Warranty Information
  • Safety Considerations
  • Unpacking and Inspection
  • Assembling the System
  • Leveling the Table
  • Principles of Operation
  • Maintenance
  • Troubleshooting
  • Factory Service

Specifications

Vertical Resonant Frequency<1.1 Hz at 80 psi
Horizontal Resonant Frequency<1.5 Hz
Recommended Load Range660 to 8,000 lb (300 to 3,600 kg)
Automatic Leveling Accuracy±0.010 inch (0.25 mm)
Vertical Adjustment Range1.3 inches (33 mm)
Typical Air Pressure Range10 to 85 psi

Frequently asked questions

What is the warranty period for Newport isolator products?

Newport warrants isolator products to be free of defects in material and workmanship for a period of two years from the date of shipment.

What safety warning is given regarding heavy payloads?

Caution must be observed to ensure that heavy payloads are not located off center on the table surface, which could cause overloading of one isolator.

What is the settling time of the system after weight removal?

The settling time after a 5-lb weight removal is less than 1.5 seconds.