This document is an instruction manual for the vibration control system. It introduces the basic information, usage methods and safety precautions of the system.
The Newport Vibration Control System provides an ideal working platform for vibration-sensitive devices such as interferometers, microscopes, and balances. The system integrates Newport's rigid, laminated honeycomb panel technology and pneumatic isolation systems to provide a mounting platform which is rigid, yet thin and light weight.
The document covers the complete lifecycle of the system, including unpacking and inspection, assembly of the table and isolators, leveling procedures, and operation principles. It also includes troubleshooting guides for common issues like poor isolation performance or air leaks, as well as maintenance and factory service information.
| Vertical Resonant Frequency | <1.1 Hz at 80 psi |
|---|---|
| Horizontal Resonant Frequency | <1.5 Hz |
| Recommended Load Range | 660 to 8,000 lb (300 to 3,600 kg) |
| Automatic Leveling Accuracy | ±0.010 inch (0.25 mm) |
| Vertical Adjustment Range | 1.3 inches (33 mm) |
| Typical Air Pressure Range | 10 to 85 psi |
Newport warrants isolator products to be free of defects in material and workmanship for a period of two years from the date of shipment.
Caution must be observed to ensure that heavy payloads are not located off center on the table surface, which could cause overloading of one isolator.
The settling time after a 5-lb weight removal is less than 1.5 seconds.