This manual is the owner's guide for Undercounter & Mid-Size Control Room Temperature with Temp Display System, which introduces the product's receiving, installation, operation, troubleshooting, maintenance and cleaning.
This manual provides instructions for installing, operating, and maintaining the Thermo Fisher Scientific 21658 undercounter and mid-size temperature control system. It includes safety guidelines, electrical requirements, setup procedures, and temperature control settings. The system is designed for scientific and medical applications, with features like a digital microprocessor controller, alarm functions, and adjustable temperature settings. Maintenance and troubleshooting steps are also covered.
The manual emphasizes safety precautions, such as avoiding storage of flammable or corrosive materials, proper grounding, and correct electrical connections. It also includes detailed steps for leveling the unit, installing wire shelves, and installing field monitor probes. Temperature settings are factory-set for optimal performance, and users are advised to consult technical support before making adjustments.
| Supply Voltage | Must conform to NEC (National Electrical Code) |
|---|---|
| Temperature Range | Factory set to operate at ultimate temperature performance |
| Set Point | 21°C or 70°F (cabinet) |
The unit is factory-set to operate at its ultimate temperature performance. Adjusting settings may alter these factory settings.
Do not store unsealed chemical materials, gasoline, or flammable liquids in the cabinet. Do not operate in the presence of explosive fumes. Ensure proper grounding and avoid using GFI circuits or extension cords.
Set a leveler on each corner of the top. Adjust the leveling legs or casters if the unit is not level. It is recommended to tilt the unit slightly toward the back (about 5 degrees) to ensure proper door closure and condensate drainage.
For visible damage, note it on the delivery receipt and take pictures. For concealed damage, stop unpacking, retain all packing materials, take pictures, and contact the carrier immediately. Do not return the unit to the manufacturer.