Quick Start Guide, provides basic operating instructions for Rosemount 1199 Sealing Systems, including operating and maintenance considerations.
This Quick Start Guide covers general handling and mechanical installation instructions for Rosemount 1199 sealing systems used with differential pressure level transmitters. It explains the function of remote seal systems, including the separation of the process fluid from the sensor using a thin flexible diaphragm and secondary fill fluid. The document also details the pressure transfer path from the process to the transmitter sensor. This manual is written in Spanish.
| Maximum static pressure limit for vacuum applications | 25 mmHgA (0.5 psia) |
|---|---|
| Minimum bend radius for capillary tubing | 8 cm (3 in.) |
| Recommended mounting height for vacuum applications | 1 m (3 ft.) below the bottom tap |
| Maximum operating temperature for PVC coating | 100 °C (212 °F) |
| Torque specification for 2500 PSIG Carbon Steel bolts | 23 ft-lb |
| Torque specification for 5000 PSIG Carbon Steel bolts | 53 ft-lb |
| Torque specification for 5000 PSIG Stainless Steel bolts | 50 ft-lb |
| Torque specification for 10000 PSIG Carbon Steel bolts | 105 ft-lb |
The maximum static pressure limit is 25 mmHgA (0.5 psia) to ensure the fill fluid remains in the liquid phase.
The minimum bend radius for the capillary tubing is 8 cm (3 in.).
Yes, a general rule is to mount the transmitter approximately 1 m (3 ft.) below the bottom tap of the vessel.
No, the PVC coating will break at temperatures near 100 °C (212 °F).