This file is the manual of FAFNIR VPI measuring sensor
This document provides instructions for the FAFNIR VPI measuring sensor and the optional VPI-Supply power adapter. It details the application scope for intrinsically safe electronic level sensors, including safety norms and installation requirements. The manual explains how to connect sensors to the transducer via a four-wire cable and how to utilize the potential equalization switch located on the circuit board. This manual is written in Portuguese.
| Auxiliary Voltage (VPI-Supply) | 24 VAC, 115 VAC or 230 VAC |
|---|---|
| Output Voltage (VPI) | 12 VDC |
| Output Current | 41.0 mA |
| Output Power | 99.8 mW |
| Inductance | 5 mH |
| Capacitance | 610 nF |
| Operating Temperature | -20 °C to +60 °C |
| Protection Rating | IP20 |
The maximum auxiliary voltage is 253 V at 230 VAC input.
Two VPI transducers can be connected to a single VPI-Supply power unit.
A minimum gap of 50 mm (arc distance) must be maintained between the two units.