This manual introduces the basic information of EUTECH fluoride ion electrode, including usage, precautions, measurement methods, etc.
The Eutech Fluoride Ion Electrode (Code no. EC-FO-03) is designed to measure fluoride ions in aqueous solutions quickly and accurately. The document outlines the required equipment, including a pH/mV meter or ion meter, and specific solutions such as Eutech Fluoride Standard Solutions and Total Ionic Strength Adjuster Buffers (TISAB 1 and TISAB 2). Proper preparation of the electrode, including removing rubber caps and filling the reference electrode, is essential before use.
| Electrode Code | EC-FO-03 |
|---|---|
| Standard Solution Concentration | 0.1 M Na F |
| Standard Solution Concentration | 1,000 ppm F-1 |
| Standard Solution Concentration | 100 ppm F-1 |
| TISAB 1 Code | EC-ISA-FL1-BT |
| TISAB 2 Code | EC-ISA-FL2-BT |
| Recommended pH Range | 5.0-5.5 |
| Temperature Limit | below 80 oC |
The electrode is operating correctly if the millivolt potential changes by 57±2 mV when the concentration changes by a factor of 10, assuming the solution temperature is between 20o and 25oC.
Correct electrode operation is indicated by a slope of 90-100% when using an ion meter.
Lab-ware made of plastic, not glass, should be used to avoid dissolving the epoxy electrode body.
Constant, but not violent, stirring is necessary for accurate measurement.