Eutech pH 5/6 & Ion 5/6 pH/mV & Ion/pH Meter User Manual

This is an instruction manual that explains the use of the pH 5/6 and Ion 5/6 meters.


Eutech pH 5/6 & Ion 5/6 pH/mV & Ion/pH Meter User Manual - cover page
Brand
EUTECH
Category
Health Care
Document type
User Manual
Language
English
Pages
27
File format
PDF
File size
234 KB
Published
18 October, 2023
Updated
18 October, 2023
Models
pH 5, Ion 5
MD5 checksum
FA7BCEDFA98FFB69EE5CF4080FBE6BDF

About this manual

This manual provides step-by-step instructions for operating the Eutech pH 5/6 and Ion 5/6 meters, including calibration procedures using various buffer standards and temperature compensation. It also covers electrode maintenance, troubleshooting, and technical specifications for each model. The meters are designed for pH, mV, and ion concentration measurements with built-in auto-buffer recognition and hold functions.

What's inside

  • Introduction
  • Getting Started
  • Calibration
  • Measurement
  • Electrode Care and Maintenance
  • Troubleshooting
  • Specifications
  • Accessories
  • Warranty
  • Return of Items

Specifications

pH Range0.00 to 14.00 pH
pH Resolution0.01 pH
pH Accuracy+/- 0.01 pH
Temperature Range0.0 to 100.0 oC
Temperature Resolution0.1 oC
Temperature Accuracy+/- 0.5 oC
Millivolt Range (pH 6)-1000 to +1000 mV
Millivolt Resolution (pH 6)1 mV
Millivolt Accuracy (pH 6)+/- 2 mV
Operating Temperature0 to 50 oC
Battery Life500 hours
Meter Dimensions14 x 7 x 3.5 cm

Frequently asked questions

What buffer standards can be used for pH calibration?

The meter supports USA (pH 4.01, 7.00, 10.01), NIST (pH 4.01, 6.86, 9.18), and Pb (pH 4.10, 6.97) buffer standards.

How do I reset the calibration values?

Press and hold the CAL key while switching on the meter, then press ENTER to confirm reset.

What is the auto-buffer recognition feature?

The meter automatically identifies the correct pH buffer values during calibration using USA, NIST, or Pb standards.

How do I condition the pH electrode before use?

Soak the electrode in pH 4 buffer solution for at least 1 hour, then rinse with tap water before calibration.

Download PDF (234 KB)