This is an instruction manual that explains the use of the pH 5/6 and Ion 5/6 meters.
This manual provides step-by-step instructions for operating the Eutech pH 5/6 and Ion 5/6 meters, including calibration procedures using various buffer standards and temperature compensation. It also covers electrode maintenance, troubleshooting, and technical specifications for each model. The meters are designed for pH, mV, and ion concentration measurements with built-in auto-buffer recognition and hold functions.
| pH Range | 0.00 to 14.00 pH |
|---|---|
| pH Resolution | 0.01 pH |
| pH Accuracy | +/- 0.01 pH |
| Temperature Range | 0.0 to 100.0 oC |
| Temperature Resolution | 0.1 oC |
| Temperature Accuracy | +/- 0.5 oC |
| Millivolt Range (pH 6) | -1000 to +1000 mV |
| Millivolt Resolution (pH 6) | 1 mV |
| Millivolt Accuracy (pH 6) | +/- 2 mV |
| Operating Temperature | 0 to 50 oC |
| Battery Life | 500 hours |
| Meter Dimensions | 14 x 7 x 3.5 cm |
The meter supports USA (pH 4.01, 7.00, 10.01), NIST (pH 4.01, 6.86, 9.18), and Pb (pH 4.10, 6.97) buffer standards.
Press and hold the CAL key while switching on the meter, then press ENTER to confirm reset.
The meter automatically identifies the correct pH buffer values during calibration using USA, NIST, or Pb standards.
Soak the electrode in pH 4 buffer solution for at least 1 hour, then rinse with tap water before calibration.