This document is an operating instructions manual for 2-wire pH or ORP Transmitter, introduces correct installation and operation methods for the product.
This manual provides operating instructions for the Eutech XTR Transmitter, a device used for the continuous measurement of pH, ORP (REDOX), Conductivity, Dissolved Oxygen, or Temperature. The transmitter outputs data via a 4-20 mA current signal and is housed in an IP65 enclosure suitable for field mounting. It is designed to operate in conjunction with appropriate sensors for various chemical and physical parameters.
| Output | 4-20 mA |
|---|---|
| Housing Protection | IP65 |
| Operating Voltage | DC 12 V to 24 V |
| Load | 100 Ohms max. for 12 V; 600 Ohms max. for 24V |
| Housing Dimensions | 77 x 77 x 28 mm |
The Amp meter should read 12mA when the electrode is immersed in a fresh pH 7.0 buffer solution.
In a pH 4.01 buffer solution, the output is 8.54mA.
In a pH 10.01 buffer solution, the output is 15.44mA.