This document is the instruction manual for the Endevco IM7264C type micro-piezoresistive accelerometer
This manual provides instructions for installing, testing, and maintaining the Endevco IM7264C piezoresistive accelerometer. It details required mounting materials, torque specifications, and precautions to avoid damage. Initial checks include impedance, insulation resistance, and zero measurand output tests. Recalibration is recommended every 6 to 12 months depending on usage. Cleaning instructions and contact information for technical support are also included.
The maximum specified mounting torque is 3.0 in-lbf (0.34 N-m).
All insulation resistance readings must be at least 100 megohms.
Recalibration should be performed at 6 to 12 month intervals.