Endevco Model 727 Piezoresistive Drop Test Accelerometers Instruction Manual

This is an instruction manual for Endevco® Model 727 piezoelectric drop test accelerometers. It introduces the features and usage of the product.


Endevco Model 727 Piezoresistive Drop Test Accelerometers Instruction Manual - cover page
Brand
Endevco
Category
Other Electronics
Document type
User Manual
Language
English
Pages
9
File format
PDF
File size
340 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
8E95431B4D9BA097D770113260F95E42

About this manual

The Endevco Model 727 is a family of rugged undamped piezoresistive accelerometers offered in a variety of ranges. It is designed for shock measurements in mobile consumer electronic devices. The manual provides technical details for the single-axis accelerometers in wax mount configuration, including handling precautions and mounting procedures.

What's inside

  • Handling Precautions
  • Initial Checkout
  • Mounting Surface Preparation
  • Cable Handling
  • Electrical Considerations
  • Excitation Voltage
  • Power and Signal Leads
  • Cable Length Considerations

Specifications

Product TypeUndamped Piezoresistive Accelerometers
ApplicationShock measurements in mobile consumer electronic devices
Mounting ConfigurationWax mount, single axis
Calibration Excitation Voltage10.000±0.005 Vdc
Maximum Excitation Voltage12 Vdc
Standard Cable Length120 inches
Input Resistance (Typical)650Ω ± 300Ω

Frequently asked questions

What precautions should be taken when handling undamped accelerometers?

Undamped accelerometers are more susceptible to damage by handling, especially unmounted units. They should be handled by properly grounded technicians at ESD safe work stations to avoid high-frequency impacts that could excite resonance.

What is the recommended adhesive for mounting the Model 727?

Aremco CrystalbondTM 509 is recommended for use up to 60°C. Cyanoacrylate (super glue) is not recommended unless an appropriate solvent is used to weaken the glue joint before removal.

How should the cable be routed to prevent damage?

The mounting preparation should allow the cable to be routed perpendicular to the primary shock direction to reduce tensile stress. The cable should be strain relieved close to the accelerometer and encapsulated with silicone RTV.