Fisher 3660 and 3661 Positioners User Manual

Fisher 3660 and 3661 Positioners: Installation Instructions


Fisher 3660 and 3661 Positioners User Manual - cover page
Brand
Fisher
Document type
User Manual
Language
Spanish
Pages
44
File format
PDF
File size
800 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
6551FCB3C247ACA8AA774A56FB29E372

About this manual

This manual provides instructions for installing, calibrating, operating, and maintaining Fisher 3660 and 3661 positioners. It includes details on mounting configurations, electrical and pneumatic connections, calibration procedures, and maintenance tasks. Safety precautions and compatibility information for actuators are also covered.

The 3660 is a pneumatic positioner, while the 3661 is an electro-pneumatic model. Both are designed for use with Fisher actuators such as 657, 667, 1250, 1250R, 3024S, and GX, and can also be mounted on Baumann actuators. The manual includes technical specifications, performance data, and safety classifications for hazardous areas. This manual is written in Spanish.

What's inside

  • Introduction
  • Scope of Manual
  • Description
  • Specifications
  • Educational Services
  • Installation
  • Hazardous Area Classifications and Safety Instructions
  • Mounting
  • Pressure Connections
  • Power Supply Connection

Specifications

Signal Input3660: 0.2–1.0 bar (3–15 psig), 0.4–2.0 bar (6–30 psig), or split range; 3661: 4–20 mA DC constant voltage up to 30 V DC
Output SignalPneumatic pressure required by the actuator, up to total supply pressure
Supply PressureRecommended: 10% above actuator requirements; Maximum: 6.2 bar (90 psig) or actuator nominal pressure, whichever is lower
FluidAir; not compatible with natural gas
Linearity±1% of output span
Hysteresis0.5% of output span
Dead Band0.1% of input span
Operating Temperature Range-40 to 82 °C (-40 to 180 °F)
Weight3660: 1.2 kg (2.6 lb); 3661: 1.4 kg (3.0 lb)

Frequently asked questions

What actuators are compatible with Fisher 3660 and 3661 positioners?

The 3660 and 3661 positioners are compatible with Fisher actuators 657, 667, 1250, 1250R, 3024S, and GX. They can also be mounted on Baumann actuators.

What is the signal input for the 3660 and 3661 positioners?

The 3660 accepts pneumatic signals (0.2–1.0 bar or 0.4–2.0 bar) or split range. The 3661 accepts a 4–20 mA DC signal with a maximum voltage of 30 V DC.

What is the operating temperature range for these positioners?

The operating temperature range is -40 to 82 °C (-40 to 180 °F).