MiniAmp™ and MiniAmp™ Plus Thermal Cycler INSTALLATION, USE, AND MAINTENANCE for use with Invitrogen™ and Applied Biosystems™ PCR reagents
This manual provides step-by-step instructions for setting up, configuring, and operating the MiniAmp™ and MiniAmp™ Plus thermal cyclers. It covers touchscreen controls, method creation, sample loading, run monitoring, and maintenance tasks like cleaning and firmware upgrades. The device is designed for PCR amplification and includes features such as remote monitoring and thermal simulation modes.
The guide also details technical specifications, environmental requirements, safety information, and troubleshooting. It is intended for users working with Invitrogen™ and Applied Biosystems™ PCR reagents and includes appendices for ordering, settings, and fleet control software.
| Maximum block ramp rate | 3.0°C/sec (MiniAmp), 3.5°C/sec (MiniAmp Plus) |
|---|---|
| Maximum sample ramp rate | 2.2°C/sec (MiniAmp), 2.7°C/sec (MiniAmp Plus) |
| Temperature accuracy | ±0.25°C (35–99.9°C) |
| Temperature range for protocol run | 0.0–100.0°C |
| Temperature non-uniformity | ≤0.5°C |
| Dimensions | Height: 20 cm (7.87 in), Width: 19 cm (7.48 in), Depth: 39 cm (15.35 in) |
| Weight | 6.75 kg (14.9 lb) |
| PCR volume range | Supported: 10–100 µL, Allowable: 1–100 µL |
| Power | 100–240 V, 50–60 Hz, Maximum 500 W |
| Display | 5 inch color TFT LCD |
The maximum block ramp rate for the MiniAmp™ is 3.0°C/sec.
Use compressed air to clear out each well of the sample block to remove particles that may have collected inside during shipping.
The temperature accuracy is ±0.25°C (35–99.9°C).