Thermo Fisher Scientific MiniAmp™ and MiniAmp™ Plus Thermal Cycler Installation, Use, and Maintenance

MiniAmp™ and MiniAmp™ Plus Thermal Cycler INSTALLATION, USE, AND MAINTENANCE for use with Invitrogen™ and Applied Biosystems™ PCR reagents


Thermo Fisher Scientific MiniAmp™ and MiniAmp™ Plus Thermal Cycler Installation, Use, and Maintenance - cover page
Brand
Thermo Fisher Scientific
Document type
User Manual
Language
English
Pages
88
File format
PDF
File size
7.9 MB
Published
17 October, 2023
Updated
17 October, 2023
MD5 checksum
829136048E26069A2A20BCA81EF2D714

About this manual

This manual provides step-by-step instructions for setting up, configuring, and operating the MiniAmp™ and MiniAmp™ Plus thermal cyclers. It covers touchscreen controls, method creation, sample loading, run monitoring, and maintenance tasks like cleaning and firmware upgrades. The device is designed for PCR amplification and includes features such as remote monitoring and thermal simulation modes.

The guide also details technical specifications, environmental requirements, safety information, and troubleshooting. It is intended for users working with Invitrogen™ and Applied Biosystems™ PCR reagents and includes appendices for ordering, settings, and fleet control software.

What's inside

  • Product Information
  • Start, Sign On, and Configure the Instrument
  • Connect the Instrument for Remote Monitoring
  • Methods
  • Maintenance
  • Troubleshooting
  • Instrument Specification and Layout
  • Safety
  • Documentation and Support

Specifications

Maximum block ramp rate3.0°C/sec (MiniAmp), 3.5°C/sec (MiniAmp Plus)
Maximum sample ramp rate2.2°C/sec (MiniAmp), 2.7°C/sec (MiniAmp Plus)
Temperature accuracy±0.25°C (35–99.9°C)
Temperature range for protocol run0.0–100.0°C
Temperature non-uniformity≤0.5°C
DimensionsHeight: 20 cm (7.87 in), Width: 19 cm (7.48 in), Depth: 39 cm (15.35 in)
Weight6.75 kg (14.9 lb)
PCR volume rangeSupported: 10–100 µL, Allowable: 1–100 µL
Power100–240 V, 50–60 Hz, Maximum 500 W
Display5 inch color TFT LCD

Frequently asked questions

What is the maximum block ramp rate for the MiniAmp™?

The maximum block ramp rate for the MiniAmp™ is 3.0°C/sec.

How do I clean the sample wells?

Use compressed air to clear out each well of the sample block to remove particles that may have collected inside during shipping.

What is the temperature accuracy of the instrument?

The temperature accuracy is ±0.25°C (35–99.9°C).