Transcend Micro is a CPAP machine that treats obstructive sleep apnea (OSA).
This manual guides users through setting up and operating the Transcend Micro, a compact CPAP device for treating obstructive sleep apnea. It covers connecting accessories like the AirFlex™ hose and WhisperSoft™ muffler, using the app, adjusting therapy settings, and maintaining the device. Technical specifications include weight, dimensions, pressure range, and operating conditions.
The manual also includes safety information, contraindications, troubleshooting, and part numbers for accessories and disposables. It is intended for use under physician supervision and includes warnings for specific medical conditions.
| Weight | 0.48 lbs (219.5g) |
|---|---|
| Dimensions | 3.6 in x 3.6“ x 2.4“ (9.1 cm x 9.1 cm x 6.1 cm) |
| Working Pressure Range | 4 to 20 cm H2O |
| Accuracy of Pressure Setting | ±1 cm H2O or ±10%, whichever is greater |
| Maximum System Shutdown Pressure | 30 cm H2O |
| Ramp Time Duration | 0-45 min + 25% time variance |
| Operating Temperature Range | 41 to 95°F (5 to 35°C) |
| Storage/Transport Temperature Range | -4 to 140ºF (-20 to 60ºC) |
| Operating Humidity Range | 10% to 80% relative humidity, non-condensing |
| Altitude Range | 0-8000 feet (Automatically adjusted) |
The Transcend Micro provides positive airway pressure for treatment of obstructive sleep apnea (OSA) in adults weighing over 66 pounds (30 kg).
The Transcend Micro may be contraindicated in patients with bullous lung disease, pathologically low blood pressure, pneumothorax, or pneumomediastinum.
The Transcend Micro weighs 0.48 lbs (219.5g).
The operating temperature range is 41 to 95°F (5 to 35°C).