Rosemount 499ACL-02 Total Chlorine Sensor Quick Start Guide

This document is a quick start guide for Rosemount 499ACL-02 Total Chlorine Sensor, which introduces the product's safety information, installation, wiring, calibration, maintenance, etc.


Rosemount 499ACL-02 Total Chlorine Sensor Quick Start Guide - cover page
Brand
Rosemount
Document type
Quick Start Guide
Language
English
Pages
16
File format
PDF
File size
802 KB
Published
16 October, 2023
Updated
16 October, 2023
MD5 checksum
314113EEBA6168CBA6B0F11F8AFB0F20

About this manual

The document serves as a quick start guide for the Rosemount 499ACL-02 Total Chlorine Sensor. It covers essential steps including unpacking, installation within the Rosemount TCL sample conditioning system, and wiring configurations for various transmitters such as models 1056, 56, 5081, and 1066.

What's inside

  • First steps
  • Install
  • Wire
  • Calibrate
  • Maintenance
  • Accessories

Specifications

Pressure (operating)0 to 65 psig (101 to 549 kPa abs)
Temperature (operating)32 to 122 °F (0 to 50 °C)
Process connection1-in. male national pipe thread (MNPT)
Wetted parts materialsNoryl®, Viton®, and silicone
Linearity2% (typical)

Frequently asked questions

What system is this sensor intended for?

The sensor is intended for use only in the Rosemount TCL sample conditioning system for total chlorine.

What are the maximum pressure and temperature limits?

Do not exceed a pressure of 65 psig (549 kPa abs) max. or a temperature range of 32 to 122 °F (0 to 50 °C).

How long does the sensor need to equilibrate after replenishing electrolyte?

The sensor may require several hours operating at the polarizing voltage to equilibrate.