This document is a quick start guide for Rosemount 499ACL-02 Total Chlorine Sensor, which introduces the product's safety information, installation, wiring, calibration, maintenance, etc.
The document serves as a quick start guide for the Rosemount 499ACL-02 Total Chlorine Sensor. It covers essential steps including unpacking, installation within the Rosemount TCL sample conditioning system, and wiring configurations for various transmitters such as models 1056, 56, 5081, and 1066.
| Pressure (operating) | 0 to 65 psig (101 to 549 kPa abs) |
|---|---|
| Temperature (operating) | 32 to 122 °F (0 to 50 °C) |
| Process connection | 1-in. male national pipe thread (MNPT) |
| Wetted parts materials | Noryl®, Viton®, and silicone |
| Linearity | 2% (typical) |
The sensor is intended for use only in the Rosemount TCL sample conditioning system for total chlorine.
Do not exceed a pressure of 65 psig (549 kPa abs) max. or a temperature range of 32 to 122 °F (0 to 50 °C).
The sensor may require several hours operating at the polarizing voltage to equilibrate.