A151 Laboratory Conductivity Meter User Manual
The Bante A151 is a laboratory conductivity meter designed for measuring conductivity, TDS, salinity, resistivity, and temperature. The device supports 2-pole and 4-pole electrodes with various cell constants (K=0.1, 1, or 10) and features automatic temperature compensation using linear or non-linear coefficients.
The manual details the installation of the electrode rack and probes, operation of the keypad functions including data logging and alarm limits, and configuration of the settings menu. It also covers communication via USB to a PC for data export and provides maintenance procedures for cleaning and calibrating the electrodes. This manual is written in Chinese.
| Conductivity Range | 0.01~20.00, 200.0, 2000µ S/cm, 20.00, 200.0mS/cm |
|---|---|
| Conductivity Accuracy | ± 0.5% F.S |
| TDS Range | 0.00mg/L~100.0g/L (max 200g/L) |
| Temperature Range | 0~105° C, 32~221° F |
| Data Storage Capacity | 1000 groups |
| Power Requirement | DC12V/2A |
The meter supports 2-pole and 4-pole electrodes with cell constants of K=0.1, K=1, or K=10.
Enter the settings menu by holding the power button for 3 seconds, select Alarm Limits, and choose Enable.