This is the Chinese user manual for the Bante A150 Laboratory Conductivity Meter, covering operation, settings, calibration, and maintenance.
This manual provides instructions for the Bante A150 Laboratory Conductivity Meter, a device used to measure conductivity, TDS, salinity, resistivity, and temperature. It details the instrument's features, including automatic reading, interval recording, and password protection, as well as the necessary accessories like the electrode stand and temperature probe. This manual is written in Chinese.
| Conductivity Range | 0.01~20.00, 200.0, 2000µ S/cm, 20.00, 200.0mS/cm |
|---|---|
| Conductivity Accuracy | ± 0.5% F.S |
| Resolution | 0.001, 0.01, 0.1, 1 |
| Calibration Points | 1 to 3 points |
| Calibration Solutions | 10µ S/cm, 84µ S/cm, 1413µ S/cm, 12.88mS/cm, 111.8mS/cm |
| Temperature Range | 0~105° C, 32~221° F |
| Power Requirement | DC12V/2A |
| Dimensions | 240 (L) × 220 (W) × 80 (H)mm |
The instrument supports calibration solutions of 10µ S/cm, 84µ S/cm, 1413µ S/cm, 12.88mS/cm, and 111.8mS/cm.
You can enable the Auto-Power Off function in the Settings Menu, which turns off the meter if there is no button operation for 3 hours.
The default reference temperature for conductivity measurement and calibration is 20°C or 25°C.
Insert the temperature probe into the round hole in the center of the electrode arm and connect its connector to the port labeled °C on the back panel.