Thermo Fisher Orion Star RDO Optical Dissolved Oxygen Sensor User Guide

A sensor for measuring dissolved oxygen content in water


Thermo Fisher Orion Star RDO Optical Dissolved Oxygen Sensor User Guide - cover page
Brand
Thermo Fisher Scientific
Document type
User Manual
Language
English
Pages
19
File format
PDF
File size
340 KB
Published
16 October, 2023
Updated
16 October, 2023
MD5 checksum
8B05C7EB9E41A404E599933D8C89E7B7

About this manual

The document provides instructions for preparing the RDO sensor by installing the cap and connecting it to a compatible meter. It details the calibration process using water-saturated air and zero oxygen solutions, along with specific preparation steps for these standards.

The guide also outlines maintenance procedures such as cleaning the optical window and replacing the RDO cap annually. Finally, it lists substances that may interfere with measurements or damage the sensor, providing troubleshooting advice.

What's inside

  • Introduction
  • Required Equipment
  • Important Notes
  • RDO Sensor Preparation
  • RDO Sensor Calibration
  • Calibration Sleeve Preparation
  • Zero Oxygen Solution for Zero Point Calibration
  • Stainless Steel Protective Guard
  • RDO Sensor Storage
  • Cleaning the RDO Cap

Specifications

Connector Type9 pin miniDIN connector
Cap Lifespan365 days

Frequently asked questions

What substances interfere with dissolved oxygen measurements?

Alcohols greater than 5%, hydrogen peroxide greater than 3%, sodium hypochlorite greater than 3%, gaseous sulfur dioxide, and gaseous chlorine may interfere.

Which substances do not interfere with dissolved oxygen measurements?

Carbon dioxide, ammonia, pH, sulfide, sulfate, chloride, and hydrogen sulfide do not interfere.

What substances should not be used with the RDO sensor?

Do not use the sensor in solutions containing organic solvents such as acetone, chloroform, or methylene chloride.