A sensor for measuring dissolved oxygen content in water
The document provides instructions for preparing the RDO sensor by installing the cap and connecting it to a compatible meter. It details the calibration process using water-saturated air and zero oxygen solutions, along with specific preparation steps for these standards.
The guide also outlines maintenance procedures such as cleaning the optical window and replacing the RDO cap annually. Finally, it lists substances that may interfere with measurements or damage the sensor, providing troubleshooting advice.
| Connector Type | 9 pin miniDIN connector |
|---|---|
| Cap Lifespan | 365 days |
Alcohols greater than 5%, hydrogen peroxide greater than 3%, sodium hypochlorite greater than 3%, gaseous sulfur dioxide, and gaseous chlorine may interfere.
Carbon dioxide, ammonia, pH, sulfide, sulfate, chloride, and hydrogen sulfide do not interfere.
Do not use the sensor in solutions containing organic solvents such as acetone, chloroform, or methylene chloride.