About this manual

The HI 93700 is a colorimeter designed to measure ammonia nitrogen (NH3-N) content in water, wastewater, and seawater within a range of 0.00 to 3.00 mg/L. The instrument utilizes an exclusive positive-locking system for cuvet placement and features an auto-shut-off function that activates after 10 minutes of non-use. It operates using a light-emitting diode light source at 470 nm and a silicon photocell detector, adhering to the ASTM D1426-92 Nessler method.

What's inside

  • Warranty
  • Preliminary Examination
  • Specifications
  • Required Reagents
  • Display Code Guide
  • Measurement Procedure
  • Interferences
  • Tips for Accurate Measurement
  • Battery Replacement
  • Accessories

Specifications

Range0.00 to 3.00 mg/L
Resolution0.01 mg/L
Accuracy+/-0.04 mg/L +/-4% of reading
Light SourceLight Emitting Diode @ 470 nm
Light DetectorSilicon Photocell
Environment0 to 50°C (32 to 122°F)
Auto-Shut offAfter 10' of non-use
Dimensions180 x 83 x 46 mm
Weight290 g

Frequently asked questions

What is the measurement range of the HI 93700?

The meter measures ammonia nitrogen content in the 0.00 to 3.00 mg/L (ppm) range.

What is the measurement range and resolution of the HI 93700 meter?

The instrument measures ammonia nitrogen content in water, wastewater, and seawater within a range of 0.00 to 3.00 mg/L with a resolution of 0.01 mg/L.

How long does the instrument remain on before it automatically shuts off?

The meter will turn itself off after 10 minutes of non-use.

How many drops of the first reagent are required for a standard water analysis?

You must add 4 drops of the First Reagent to perform the test.