This document introduces the OQ/PV test flow and method of Agilent Technologies liquid chromatograph 1100, the flow can test the whole system, including ChemStation software and hardware
This document outlines the On-qualification (OQ) and Performance Verification (PV) procedures available from Agilent Technologies for the Agilent 1100 Series Liquid Chromatograph. The service includes internal tests and analytical methods using certified samples to evaluate instrument performance and verify operation against specifications.
The process utilizes the ChemStation software, version A.06.03 or higher, which automates testing of the entire system in its original configuration. This approach minimizes downtime, typically allowing the full procedure to be completed within a single day while ensuring results accurately reflect the system's performance in the specific user environment. This manual is written in German.
The full OQ/PV procedure can usually be performed in a single day to minimize downtime.
The system requires ChemStation version A.06.03 or higher to perform the OQ/PV procedures.
Tests for the variable wavelength detector include lamp intensity, Holmium oxide, and wavelength accuracy.