This document is OQ/PV validation procedure for Agilent 6890 series gas chromatograph.
The document outlines the OQ/PV procedures for the Agilent Technologies GC Series 6890. These procedures include internal tests and analytical methods designed to evaluate system performance and verify specifications against user requirements and Agilent standards. The process typically takes one day, minimizing downtime while ensuring that instruments meet optimal performance criteria through maintenance prior to testing.
The guide specifies supported hardware configurations, including split/splitless injection, packed columns, on-column injection, programmable temperature vaporizers, various detectors (FID, TCD, NPD, ECD, MS), and specific autosamplers. It details test procedures for system maintenance, leak tightness, oven temperature stability, detector sensitivity, and chemical performance. This manual is written in German.
| Software Version | ChemStation Software A.05.XX or higher |
|---|---|
| Supported Autosamplers | 7673 and 7683 |
The complete procedure can usually be performed in a single day to minimize downtime.
System maintenance must be performed prior to the OQ procedure to guarantee full performance.