Agilent SecureTD Application Note for Re-collection Repeat Analysis

Introduced a new thermal desorption technology, which has the advantages of high sensitivity, high recovery rate, no solvent and low cost.


Agilent SecureTD Application Note for Re-collection Repeat Analysis - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
329 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
F71A2BAA5360CC3DCB8A4DED93D6193F

About this manual

Thermal desorption (TD) offers significant advantages over solvent extraction, including increased sensitivity, higher sample recovery, and cost savings. However, conventional TD is a one-shot process where the sample is destroyed during analysis. Agilent's SecureTD technology, available on Markes International TD systems like UNITY and ULTRA-UNITY, overcomes this limitation by using split re-collection to capture a portion of the desorbed sample for repeat analysis.

The document details the technical overview of SecureTD, explaining how split flows are directed to charcoal or conditioned sorbent tubes for quantitative trapping. It presents laboratory test results demonstrating the performance of SecureTD with high-loading toluene standards and low-concentration benzene, toluene, and o-xylene solutions. The note also covers method validation for high-boiling compounds and troubleshooting scenarios where repeat analysis helps identify issues such as erroneous GC parameters or sample loading problems.

What's inside

  • Introduction
  • Technical Overview
  • SecureTD Split Re-collection
  • Laboratory Test of Re-collection Performance
  • Method Validation
  • Erroneous GC Parameters
  • Troubleshooting

Specifications

Split ratio (Toluene test)99:1
Split ratio (BTX test)21:1
Split ratio (Amines test)~130:1
Tube desorb temperature (Toluene)280 °C for 5 min
Tube desorb temperature (Amines)320 °C for 10 min

Frequently asked questions

What is the main advantage of SecureTD over conventional thermal desorption?

SecureTD overcomes the one-shot limitation of conventional TD by using quantitative re-collection for repeat analysis.

Which Markes International systems support SecureTD?

Markes International TD systems, including UNITY (single tube thermal desorber) and ULTRA-UNITY (100-tube automated thermal desorber), support SecureTD.