Agilent PL Rapide Aqua Aqueous SEC Columns Datasheet

This document introduces the installation steps of Agilent PL Rapide Aqua Rapid Aqueous SEC Columns, including connecting pipes and connectors, column connections, etc.


Agilent PL Rapide Aqua Aqueous SEC Columns Datasheet - cover page
Brand
Agilent
Document type
Datasheet
Language
English
Pages
2
File format
PDF
File size
341 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
9B1AA27965A63C1C1E796A64147747D5

About this manual

This document provides technical data and operational guidelines for the Agilent PL Rapide Aqua series of aqueous SEC columns. It details recommended tubing connections, compatible fittings, and specific procedures for column installation to ensure optimal system performance.

The guide covers critical operating parameters such as eluent selection (including pH limits and organic solvent restrictions), maximum pressure and temperature ratings, and flow rate recommendations based on column dimensions. It also includes instructions for sample preparation, injection volumes, and routine maintenance like reverse flushing frit blockages.

What's inside

  • Tubing and connectors
  • Column connection
  • Eluent Selection and Flow Rate
  • Sample Preparation and Injection
  • Column Testing and Specifications
  • Storage
  • Maintenance

Specifications

Typical operating pressure225 psi (15 bar)
Maximum operating pressure2000 psi (140 bar)
Maximum operating temperature40 °C
Efficiency (H2)> 35,000 plates/m

Frequently asked questions

What organic solvent is recommended for use with these columns?

The only recommended organic solvent is methanol, up to 50% by volume.

What is the maximum pH range for buffered eluent systems?

Buffered eluent systems within the pH range of 2-10 may be used.

How should a blocked column frit be maintained?

The column can be reverse flushed at 1.0 mL/min for 1 minute to remove loosely retained material.