The Agilent 8000 Dissolution Sampling Station provides automated sampling for our industry-leading dissolution apparatus. Designed to be configured with either the 810 Peristaltic Pump, or 806 Syringe Pump and 808 Filter Changer, the integrated system efficiently provides accurate and precise sampling at each timepoint.
The Agilent 8000 Dissolution Sampling Station is designed to automate sampling for dissolution testing, eliminating the inaccuracies associated with manual methods. It can be configured with either the 810 Peristaltic Pump or the 806 Syringe Pump and 808 Filter Changer to accommodate various sampling needs, from standard test tubes to HPLC vials.
The system features automated calibration, a built-in printer for comprehensive test data, and the ability to operate with both Agilent and non-Agilent dissolution apparatus. It supports up to 20 timepoints per run and includes options for automated media replacement to maintain constant dissolution volumes.
| Altitude | 0 - 2000 m (0 - 6562 ft.) |
|---|---|
| Temperature | 5 to 40 °C |
| Humidity | Not more than 80% RH |
| Voltage requirements | 115 V/ 60 Hz or 230 V/ 50 Hz |
| Current requirements | 2.0 Amp |
| Fuse requirements | 2.0 Amp, 250 V, 5 mm x 20 mm FAST |
| Sampling accuracy (810 Peristaltic Pump) | + 0.5 mL |
| Sampling accuracy (806 Syringe Pump) | + 0.2 mL |
| Sampling accuracy (806 Syringe Pump + 808 Filter Changer) | + 0.3 mL |
| Number of sample points / run | 1 - 10; up to 20 with manual sample tray exchange |
| Maximum run time | 999 hours |
| Sample volumes | 1 - 14mL |
| Program storage | 15 programs with up to 10 sample timepoints each |
| Dimensions (Height) | 41.28 cm (16.25 in.) |
| Dimensions (Width) | 43.18 cm (17 in.) |
| Dimensions (Depth) | 51.8 cm (20.38 in.) |
| Equipment weight | 16.3 kg (36 lb) |
The 8000 may be programmed to operate with either the 810 Peristaltic Pump or the 806 Syringe Pump.
Yes, the system can be programmed to operate any of Agilent's dissolution apparatus as well as several non-Agilent dissolution apparatus.
Yes, it can be equipped with an optional media replacement module to replace sample volumes removed during the run.
Yes, Agilent's UV Dissolution Software and/or Dissolution Workstation Software enable compliance with 21 CFR Part 11 guidelines.