Agilent SPS-5 Autosampler Application Note for Quinine Sulfate Analysis

This application note describes an automated analysis method for determining the uniformity of dosage in quinine sulfate tablets using an Agilent fiber-optic autosampler. The method uses a Cary 3C double beam UV/Vis spectrophotometer, Agilent SPS-5 autosampler, Cary fiber probe, and Cary fiber optic coupler. A Cary OS2 concentration application module and simple ADL program are used to locate the peak maximum and its absorbance value, from which the concentration of the analyte can be determined from known standards or a known extinction coefficient. This method has been applied to the measurement of uniformity of dosage for quinine sulfate in tablets using the USP procedure.


Agilent SPS-5 Autosampler Application Note for Quinine Sulfate Analysis - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
199 KB
Published
11 June, 2012
Updated
29 September, 2023
Model
SPS-5
MD5 checksum
0848449F7E51B05E437BE5EF6D2440B0

About this manual

This document details an analytical procedure for measuring the uniformity of dosage in quinine sulfate tablets. It utilizes the Agilent SPS-5 Autosampler coupled with a Cary 3C Double Beam UV/Vis Spectrophotometer and the Cary 082 Concentration Application module. The method employs a specific ADL program to locate peak maximum absorbance within a defined range, ensuring accurate concentration determination from known standards or extinction coefficients.

What's inside

  • Introduction
  • Experimental
  • Apparatus
  • Reagents
  • ADL Program
  • Procedure
  • Standard Preparation
  • Sample Preparation
  • Analysis Procedure
  • Results

Specifications

Wavelength scan range400 to 320 nm
Correlation coefficient0.99998
Standard deviation0.02

Frequently asked questions

What software is used for the concentration analysis?

The Cary 082 Concentration Application module with Cary OS2 software is used.

What is the allowable wavelength range for the peak maximum?

The peak maximum must occur between 320 nm and 360 nm.

How many tablets are selected for the sample preparation?

Ten tablets are selected and transferred to individual volumetric flasks.