Varian Packed GC Column User Guide

This document provides a user guide for Varian Packed GC Column, including instructions for preconditioning, gas purification, and optimum flow.


Varian Packed GC Column User Guide - cover page
Brand
Varian
Document type
User Manual
Language
English
Pages
1
File format
PDF
File size
70 KB
Published
07 June, 2012
Updated
28 September, 2023
MD5 checksum
2CB0BB372C1DD311E49C993089802E42

About this manual

The document outlines a procedure for preconditioning Varian packed GC columns. It specifies heating temperatures relative to the maximum operating temperature (20 °C below) and provides ramp rates from 30 °C to 300 °C at 0.5 °C/min depending on column dimensions.

Additionally, the guide emphasizes the importance of using purified carrier gas with a minimum flow of 20 mL/min. It lists specific Varian Gas Clean filters (Oxygen, Moisture, Charcoal, and GC/MS) designed to prevent contamination that could cause damage or loss of sensitivity.

What's inside

  • Preconditioning
  • Heat the column
  • Gas Purification
  • Ordering Information

Frequently asked questions

What is the maximum temperature for preconditioning the column?

The column should be heated to 20 °C below its maximum operating temperature.

Download PDF (70 KB)

More Varian Manuals


View all Varian manuals →