This document introduces information about Agilent Max-Light Cartridge Cell product, including product specifications, installation and use and cleaning and maintenance
This document provides guidance on the use, installation, storage, and cleaning of Agilent Max-Light Cartridge Cells for the Agilent 1290/1260 Infinity Diode Array Detector. It details specifications for various flow cell models, including standard, bio-inert, HDR, and ULD versions.
The guide covers specific application recommendations for protein analysis and high pH mobile phases, installation procedures to prevent contamination, and storage instructions using iso-propanol or methanol. It also outlines firmware requirements for the 60 mm flow cells.
| Maximum pressure | 60 bar (870 psi) |
|---|---|
| pH range | 1.0-12.5 |
| Dispersion volume V(o) for G4212-60007 | 4.0 pL |
| Dispersion volume V(o) for G4212-60008 | 1.0 pL |
| Required firmware for 60 mm cell | B.06.26 or later |
The usage of an Agilent Diode Array Detector G1315C/D with the respective bio-inert flow cell is highly recommended.
Flush the cell with iso-propanol or methanol, insert plugs, remove from the drawer, replace hoods, and store in a safe location.
Users should not let buffers stay for long times; they must flush the Max-Light Cartridge Flow Cell when finished with the application.