Agilent Technologies Max-Light Cartridge Cell Information for G4212A G4212B DAD guide

This document introduces information about Agilent Max-Light Cartridge Cell product, including product specifications, installation and use and cleaning and maintenance


Agilent Technologies Max-Light Cartridge Cell Information for G4212A G4212B DAD guide - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
8
File format
PDF
File size
185 KB
Published
06 June, 2012
Updated
30 September, 2023
MD5 checksum
F0CBA6FF5BBC716233FF09DD3A54580C

About this manual

This document provides guidance on the use, installation, storage, and cleaning of Agilent Max-Light Cartridge Cells for the Agilent 1290/1260 Infinity Diode Array Detector. It details specifications for various flow cell models, including standard, bio-inert, HDR, and ULD versions.

The guide covers specific application recommendations for protein analysis and high pH mobile phases, installation procedures to prevent contamination, and storage instructions using iso-propanol or methanol. It also outlines firmware requirements for the 60 mm flow cells.

What's inside

  • Specifications
  • Description of dispersion volume
  • Recommendations
  • Application Information
  • Installation
  • During Use
  • Storage
  • Cleaning
  • Special Information for 60 mm Cell
  • Bio-Inert Cartridge Flow Cells

Specifications

Maximum pressure60 bar (870 psi)
pH range1.0-12.5
Dispersion volume V(o) for G4212-600074.0 pL
Dispersion volume V(o) for G4212-600081.0 pL
Required firmware for 60 mm cellB.06.26 or later

Frequently asked questions

What is recommended for cation exchange chromatography to avoid protein interaction?

The usage of an Agilent Diode Array Detector G1315C/D with the respective bio-inert flow cell is highly recommended.

How should Max-Light Cartridge Cells be stored?

Flush the cell with iso-propanol or methanol, insert plugs, remove from the drawer, replace hoods, and store in a safe location.

What happens if buffers stay in the flow cell for long times?

Users should not let buffers stay for long times; they must flush the Max-Light Cartridge Flow Cell when finished with the application.