This document provides notices and information about the performance report parameters of Agilent MSD Productivity ChemStation G1701 & G1710, including the prohibition of unauthorized reproduction and translation, disclaimer, and warranty information.
This document describes the specific calculations used by the Performance Report feature in the Agilent MSD Productivity ChemStation. The report generation depends on the integrator selected, supporting both the standard ChemStation integrator and the RTE (Retention Time Editor) integrator.
The guide details how to generate the report via the Chromatogram menu and defines the structure of the output, including the Heading section, Graphics Area, Noise characteristics, and Performance Report Results table.
For the ChemStation Integrator, the peak width is defined as the width of the peak at 50% of the peak height.
The RTE Integrator calculates resolution using the peak width at the baseline instead of the peak width at 50% height.
The user must enter a specific retention time range where the software will calculate the noise.