Agilent MSD Productivity ChemStation G1701 Performance Report Parameters Guide

This document provides notices and information about the performance report parameters of Agilent MSD Productivity ChemStation G1701 & G1710, including the prohibition of unauthorized reproduction and translation, disclaimer, and warranty information.


Agilent MSD Productivity ChemStation G1701 Performance Report Parameters Guide - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
10
File format
PDF
File size
707 KB
Published
06 June, 2012
Updated
29 September, 2023
Model
G1701
MD5 checksum
6563FC86861EE4741783FDB205CEDEA5

About this manual

This document describes the specific calculations used by the Performance Report feature in the Agilent MSD Productivity ChemStation. The report generation depends on the integrator selected, supporting both the standard ChemStation integrator and the RTE (Retention Time Editor) integrator.

The guide details how to generate the report via the Chromatogram menu and defines the structure of the output, including the Heading section, Graphics Area, Noise characteristics, and Performance Report Results table.

What's inside

  • Introduction
  • Report Layout
  • Heading
  • RTE Integration Parameters
  • Graphics Area
  • Noise characteristics
  • Performance Report Results

Frequently asked questions

How is the peak width calculated for the ChemStation Integrator?

For the ChemStation Integrator, the peak width is defined as the width of the peak at 50% of the peak height.

How does the RTE Integrator calculate resolution differently from the ChemStation Integrator?

The RTE Integrator calculates resolution using the peak width at the baseline instead of the peak width at 50% height.

What retention time range must a user specify to calculate noise?

The user must enter a specific retention time range where the software will calculate the noise.

Download PDF (707 KB)

More Agilent Manuals


View all Agilent manuals →