Yokogawa ZR22G/ZR402G/ZR202G Direct In Situ Zirconia Oxygen and High Temperature Humidity Analyzer User Manual

This document is about Yokogawa Electric Corporation's ZR22G, ZR402G, ZR202G type direct in situ zirconia oxygen analyzers and high temperature humidity analyzers.


Yokogawa ZR22G/ZR402G/ZR202G Direct In Situ Zirconia Oxygen and High Temperature Humidity Analyzer User Manual - cover page
Brand
YOKOGAWA
Category
Industrial Supply
Document type
User Manual
Language
English
Pages
36
File format
PDF
File size
711 KB
Published
05 January, 2012
Updated
30 September, 2023
Models
ZR202G, ZR402G
MD5 checksum
EBF8A5B5739575F01F6FCA6E7A741D42

About this manual

The manual describes Yokogawa's direct‑in‑situ zirconia oxygen analyzers and high‑temperature humidity analyzers, available in separate and integrated configurations. It covers system components, installation, calibration, operation, and maintenance procedures.

Key features include a digital LCD touchscreen, HART digital communication, remote maintenance, replaceable probe heater, and multiple reference‑air supply options. The document provides detailed specifications, wiring guidelines, and safety standards for reliable use in boilers, furnaces, and drying equipment.

What's inside

  • Overview
  • Basic System Configuration
  • System Components
  • Standard Specifications
  • Oxygen Analyzer Specifications
  • Calibration Functions
  • Display and Setting Content
  • Installation and Wiring
  • Safety and EMC Standards
  • Maintenance and Self‑diagnosis

Specifications

Measurement Range0 to 100 vol% O
Output Signal4 to 20 mA DC (max load 550 Ω)
Warm‑up TimeApprox. 20 min
Response Time90 % within 5 s
Sample Gas Temperature (probe only)0 to 700 °C
Sample Gas Temperature (with high‑temp probe)700 to 1400 °C
Probe Length Options0.15 m to 5.4 m
Ambient Temperature Range-20 °C to +150 °C
Power Supply Voltage100 V to 240 V AC (85 V to 264 V acceptable)
Power ConsumptionMax 300 W (≈100 W typical)
Maximum Probe‑to‑Converter Distance≤300 m (conductor resistance ≤100 Ω)

Frequently asked questions

How long does the analyzer take to warm up?

The warm‑up time is approximately 20 minutes.

What is the maximum distance allowed between the probe and the converter?

The conductor resistance must be 100 Ω or less, which corresponds to up to 300 m when using a 1.25 mm² cable.

Download PDF (711 KB)

More YOKOGAWA Manuals


View all YOKOGAWA manuals →