YOKOGAWA ZR402G Manuals

ManualLib has 2 documents for the YOKOGAWA ZR402G: 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
ZR402G
Documents
2
Languages
EN · JA
Last added
2012-01-05

Start here

Yokogawa ZR22G/ZR402G/ZR202G Direct In Situ Zirconia Oxygen and High Temperature Humidity Analyzer User Manual

Yokogawa ZR22G/ZR402G/ZR202G Direct In Situ Zirconia Oxygen and High Temperature Humidity Analyzer User Manual

User Manual 36 pages · 711 KB · EN

The manual describes Yokogawa's direct‑in‑situ zirconia oxygen analyzers and high‑temperature humidity analyzers, available in separate and integrated configurations. It covers system components, installation, calibration, operation, and maintenance procedures.

Key features include a digital LCD touchscreen, HART digital communication, remote maintenance, replaceable probe heater, and multiple reference‑air supply options. The document provides detailed specifications, wiring guidelines, and safety standards for reliable use in boilers, furnaces, and drying equipment.

Read online

YOKOGAWA ZR402G specifications

Measurement Range0 to 100 vol% O
Output Signal4 to 20 mA DC (max load 550 Ω)
Warm‑up TimeApprox. 20 min
Response Time90 % within 5 s
Sample Gas Temperature (probe only)0 to 700 °C
Sample Gas Temperature (with high‑temp probe)700 to 1400 °C
Probe Length Options0.15 m to 5.4 m
Ambient Temperature Range-20 °C to +150 °C
Power Supply Voltage100 V to 240 V AC (85 V to 264 V acceptable)
Power ConsumptionMax 300 W (≈100 W typical)
Maximum Probe‑to‑Converter Distance≤300 m (conductor resistance ≤100 Ω)

Source: Yokogawa ZR22G/ZR402G/ZR202G Direct In Situ Zirconia Oxygen and High Temperature Humidity Analyzer User Manual

All YOKOGAWA ZR402G documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

How long does the analyzer take to warm up?

The warm‑up time is approximately 20 minutes.

From: Yokogawa ZR22G/ZR402G/ZR202G Direct In Situ Zirconia Oxygen and High Temperature Humidity Analyzer User Manual

What is the maximum distance allowed between the probe and the converter?

The conductor resistance must be 100 Ω or less, which corresponds to up to 300 m when using a 1.25 mm² cable.

From: Yokogawa ZR22G/ZR402G/ZR202G Direct In Situ Zirconia Oxygen and High Temperature Humidity Analyzer User Manual

Can the detector be replaced without shutting down the system?

Yes, the manual states that hot‑swap of the detector is possible by powering off the channel and replacing the detector.

From: YOKOGAWA ZR202G/ZR402G AV55G Manual

How are error messages displayed to the user?

Error numbers and their descriptions are shown on the LCD screen, allowing operation without a separate instruction manual.

From: YOKOGAWA ZR202G/ZR402G AV55G Manual