Key facts
- Brand
- Simtronics
- Model
- GD10P
- Category
- sensor
- Documents
- 4
- Languages
- PT · EN
- Last added
- 2023-11-17
Start here

Simtronics GD10P Infrared Gas Detector Operations Manual
The Simtronics GD10P is an infrared point gas detector designed for effective gas risk detection in diverse industrial environments, ranging from boiler rooms to onshore petrochemical plants. It utilizes solid-state silicon infrared sources to provide fast response times and long-term stability, reducing maintenance costs.
Compared to catalytic sensors, the GD10P does not require oxygen for measurement, making it suitable for inert gas atmospheres. It is immune to contamination from silicon vapors and H2S, and it features a continuous self-test function to communicate optical fouling or errors to the control system. This manual is written in Portuguese.
Simtronics GD10P specifications
| Detection Method | IV Absorption, dual wavelength, dual path |
|---|---|
| Response Time (T90) | 6 sec (Option 1.6 sec) |
| Power Supply | 24 V DC (range 18-32 V DC) |
| Power Consumption | Approx. 3.5 W |
| Operating Temperature | -40°C to +65°C |
| Protection Category | IP66/IP67 |
| Housing Material | Stainless steel SIS2343 (ASTM 316) |
| Weight | Approx. 2.9 kg |
Source: Simtronics GD10P Infrared Gas Detector Operations Manual
All Simtronics GD10P documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
Where should the GD10P be mounted to detect methane?
Since methane is lighter than air, the detector should be mounted in the high part of the internal area or immediately above possible leak sites.
From: Simtronics GD10P Infrared Gas Detector Operations Manual
What is the warranty period for the GD10P?
The instrument has a total warranty of 5 years, while the IV sources have a 15-year warranty.
From: Simtronics GD10P Infrared Gas Detector Operations Manual
Does the GD10PE need field recalibration?
No. The GD10PE stays within its specifications for its service lifetime without recalibration; calibration is factory set.
Where should the detector be mounted to detect methane?
Inside an enclosed area it should be mounted high in the area to be protected or immediately above potential leakage sites, since methane is lighter than air.
What gases can the GD10P detect?
The detector is primarily designed for hydrocarbon or CO2 gases but detects ethane, propane, and butane as well.
Does the detector require oxygen for measurement?
No, the presence of oxygen is not required for correct measurement.