Key facts
- Brand
- Rotronic
- Model
- HygroClip2 Advanced
- Category
- sensor
- Industry
- Industrial
- Documents
- 3
- Languages
- EN
- Last added
- 2023-10-16
Start here

ROTRONIC HygroClip2 ADVANCED Humidity Temperature Probes User Manual
The HC2A series are digital humidity-temperature probes based on AirChip3000 technology. They feature a UART serial interface and two 0…1 V linear analog output signals. These probes are designed for use with the HygroClip 2 generation of ROTRONIC instruments, including transmitters, data loggers, and hand-helds, but can also function as standalone devices.
Read onlineRotronic HygroClip2 Advanced specifications
| Operating humidity | 0...100 %rh |
|---|---|
| Operating temperature | -50…100 °C |
| Accuracy | ±0.8 %rh |
| Supply voltage | 3.3…5 VDC |
| Current consumption | Approx. 5 mA |
| Cable length | Max 5 m |
| Protection rating | IP65 |
| Analog outputs | 2x 0…1 V |
| Digital communication | UART |
Source: Rotronic HygroClip2 Advanced Short Instruction Manual
All Rotronic HygroClip2 Advanced documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
When should the HygroClip2 Advanced be calibrated?
We recommend a calibration after 1 year of use, depending on the application.
Which Rotronic products is the HygroClip2 Advanced compatible with?
It is compatible with handhelds of the HP3x family and transmitters of the HF5 and HF8 families.
How often should the probes be calibrated?
We recommend a calibration after 1 year of use, depending on the application.
What tools are needed for on-site calibration?
You can perform on-site calibration using the Rotronic SCS humidity standards or the HG2-S humidity generators.
What is the minimum load requirement for probe voltage outputs?
The manual specifies a section on 'Minimum load requirements for the probe voltage outputs' but does not state the specific value in the provided text.
From: ROTRONIC HygroClip2 ADVANCED Humidity Temperature Probes User Manual
How can the HC2A probes be updated?
The ability to update the AirChip3000 firmware allows the probes to be kept up-to-date regarding future functionality improvements.
From: ROTRONIC HygroClip2 ADVANCED Humidity Temperature Probes User Manual