Rosemount 499ACL-02 Manuals

ManualLib has 2 documents for the Rosemount 499ACL-02 (Industrial Robot): 2 quick start guides. Every file can be read online or downloaded as a PDF for free.

Quick Start Guide 2

Key facts

Model
499ACL-02
Documents
2
Languages
EN
Last added
2023-10-16

Start here

Rosemount 499ACL-02 Total Chlorine Sensor Quick Start Guide

Rosemount 499ACL-02 Total Chlorine Sensor Quick Start Guide

Quick Start Guide 24 pages · 1.1 MB · EN

The Rosemount 499ACL-02 is a total chlorine sensor designed for use with the Rosemount TCL sample conditioning system. The document serves as a quick start guide, covering the initial unpacking and inspection of the sensor components, which include the membrane assembly, cathode, and electrolyte fill plug.

The guide details the installation process, noting that the sensor must be used in a specific flow cell (PN 24091-01). It also outlines procedures for zero and full-scale calibration, as well as maintenance tasks such as cleaning the membrane and replacing the electrolyte solution.

Read online

Rosemount 499ACL-02 specifications

Pressure0 to 65 psig (101 to 549 kPa abs)
Temperature (operating)0 to 50 °C (32 to 122 F)
Wetted partsNoryl, Viton, and silicone
CathodeGold
Process connectionSensor must be used in flow cell PN 24091-01

Source: Rosemount 499ACL-02 Total Chlorine Sensor Quick Start Guide

All Rosemount 499ACL-02 documents by type

Quick Start Guide 2

The shortest path to a working setup, usually only a few pages.

Questions answered in these manuals

What is the maximum operating pressure for this sensor?

The maximum operating pressure is 65 psig.

From: Rosemount 499ACL-02 Total Chlorine Sensor Quick Start Guide

What is the allowable temperature range for operation?

The sensor operates between 32 and 122 °F.

From: Rosemount 499ACL-02 Total Chlorine Sensor Quick Start Guide

What system is this sensor intended for?

The sensor is intended for use only in the Rosemount TCL sample conditioning system for total chlorine.

From: Rosemount 499ACL-02 Total Chlorine Sensor Quick Start Guide

What are the maximum pressure and temperature limits?

Do not exceed a pressure of 65 psig (549 kPa abs) max. or a temperature range of 32 to 122 °F (0 to 50 °C).

From: Rosemount 499ACL-02 Total Chlorine Sensor Quick Start Guide