Rosemount 5081-P Two-Wire pH/ORP Transmitter User Manual

Rosemount 5081-P Two-Wire pH/ORP Transmitter User Manual

User Manual 156 pages · 5.2 MB · EN

This manual covers the Rosemount 5081-P two‑wire pH/ORP transmitter, a rugged Type 4X (IP65) device for measuring pH or oxidation‑reduction potential in process liquids. It details the transmitter’s features, specifications, hazardous‑location approvals, and communication options (HART or FOUNDATION Fieldbus).\n\nThe document includes step‑by‑step procedures for installation, wiring, intrinsic‑safe and explosion‑proof setups, operation with the infrared remote controller or Model 375 HART communicator, calibration of pH and ORP, programming parameters, routine maintenance, and troubleshooting guidance.

Read online

Rosemount 5081-P specifications

HousingCast aluminum with epoxy coating, Type 4X (IP65)
Dimensions160.5 mm × 175.3 mm × 161.3 mm
Ambient Temperature Range-4 °F to 149 °F (‑20 °C to 65 °C)
Power Supply Voltage12 V DC minimum, 42.4 V DC maximum (30 V DC for intrinsically safe operation)
Analog OutputTwo‑wire 4‑20 mA with superimposed HART digital signal
pH Measurement Range0 to 14
ORP Measurement Range-1400 mV to +1400 mV
Temperature Resolution0.1 °C

Source: Rosemount 5081-P Two-Wire pH/ORP Transmitter User Manual

All Rosemount 5081-P documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the power supply voltage range for this transmitter?

The device operates on a power supply of 9 to 32 VDC.

From: Rosemount Model 5081-P pH/ORP Transmitter User Manual

What power supply voltage is required for the Model 5081-P?

The transmitter requires a DC power supply between 9 and 32 volts.

From: Rosemount Model 5081-P pH/ORP Transmitter User Manual

How many wires does the transmitter use for power and signal?

It uses a two‑wire loop delivering a 4‑20 mA current signal with a superimposed HART digital signal.

From: Rosemount 5081-P Two-Wire pH/ORP Transmitter User Manual

What hazardous‑location approvals does the transmitter have?

The transmitter has a robust Type 4X enclosure and an intrinsically safe design for use in hazardous environments when proper safety barriers are applied.

From: Rosemount 5081-P Two-Wire pH/ORP Transmitter User Manual