Hanna Instruments HI 98509-01/98510-01 Precision Thermometers Instruction Manual
This manual provides instructions for the Checktemp-Dip precision thermometers, models HI 98509-01 and HI 98510-01. These instruments feature a fixed stainless steel probe with a 3-meter cable, designed for measuring temperature at different depths in vats, tanks, wine casks, and milk tanks. The device includes a Cal-Check technology to verify measurement accuracy.
Read onlineHanna Instruments HI 98510 specifications
| Range | -20.0 to 100.0°C / -4.0 to 212°F |
|---|---|
| Resolution | 0.1°C / 0.1°F (-4 to 199.9°F) / 1°F (200 to 212°F) |
| Accuracy | +0.5°C (-20 to 50°C) / +0.9°C (outside) |
| Probe | Stainless steel, 3m (9.9) cable |
| Battery | 1x1.5V AAA alkaline |
| Battery Life | Approx. 3 years of use |
| Dimensions | 106 x 58 x 19 mm (4.1x5.2x0.7") |
| Weight | 80 g (2.8 02.) |
Source: Hanna Instruments HI 98509-01/98510-01 Precision Thermometers Instruction Manual
All Hanna Instruments HI 98510 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the ideal operating temperature range for the thermometer?
The ideal operating temperature is in the -20 to 50°C (-4 to 120°F) range.
From: Hanna Instruments HI 98509-01/98510-01 Precision Thermometers Instruction Manual
How do I check if the thermometer is measuring correctly?
Move the Cal-check switch to the 'Test' position; the display should show 0.0°C (+0.5°C) or 32°F (+0.9°F).
From: Hanna Instruments HI 98509-01/98510-01 Precision Thermometers Instruction Manual
How do I perform a calibration check?
Move the Cal-check switch on the rear of the meter to the "Test" position and wait for the display to stabilize.
From: Hanna Instruments HI 98509/98510 Pocket Thermometers Instruction Manual
What should the display show during a calibration check?
The display should show 0.0°C (+0.3°C) or 32°F (+/-0.5°F).