Druck DPI 610/615 IS Portable Pressure Calibrator User Manual
This is the German-language operating manual (Bedienungsanleitung, KD0430) for the Druck DPI 610 IS and DPI 615 IS intrinsically safe portable pressure calibrators, published by GE Sensing & Inspection Technologies. It contains safety instructions, ATEX marking and installation requirements, a full specification, and step-by-step operating procedures.
The manual describes the control panel and display, battery installation, pressure, voltage and current measurement, and the standard and advanced calibration modes (transmitter, pressure transmitter, switch test, leak test, gauge and simulation tasks). It also covers memory and data-logger operations, setup options, instrument calibration procedures, and the hydraulic calibrator versions, with appendices showing a data-logger file example and the ATEX conformity certificate. This manual is written in German.
Druck DPI 610/615 specifications
| Safe working pressure, 20 bar range | 1.75 x full-scale |
|---|---|
| Safe working pressure, 350 bar range | 1.2 x full-scale |
| Safe working pressure, 400 bar range | 1.5 x full-scale |
| Safe working pressure, all other ranges | 2 x full-scale |
| Accuracy | Combined non-linearity, hysteresis and repeatability: ±70 mbar range 0.05% F.S.; up to ±150 mbar 0.05% span; 200 mbar to 20 bar [Calibrator] 0.025% F.S.; 35 bar to 700 bar [Indicator] 0.025% F.S.; 70 bar to 400 bar [Hydraulic] 0.025% F.S. |
| Temperature effects | ±0.004% of reading/°C averaged over -10° to +40°C w.r.t. 20°C; ±0.002% of reading/°F averaged over +14° to 104°F w.r.t. 68°F |
| Power supply | 6 x 1.5 V C cells, alkaline (up to 60 hours nominal use at 20°C); rechargeable NiCad battery pack (20 hours nominal use) with charger/adaptor |
| Voltage inputs | Range ±50V; accuracy ±0.05% rdg, ±0.004% F.S.; resolution 100µV max |
| Voltage outputs | Range ±10V, accuracy ±0.1%, load 10mA; range ±24V, accuracy ±5%, load 26mA |
| Current inputs | Range ±55mA; accuracy ±0.05% rdg, ±0.004% F.S.; resolution 1µA max |
| Current output | Range 24mA; accuracy ±0.05% rdg, ±0.01% F.S.; resolution 1µA max |
| Display | 60 x 60 mm (2.36 in x 2.36 in) LCD graphics; reading ±99999; update rate 2 readings/sec |
| Operating temperature | -10°C to 50°C (+14°F to 122°F) |
| Calibrated temperature | -10°C to 40°C (+14°F to 104°F) |
| Storage temperature | -20°C to 60°C (-4°F to 140°F) |
| Calibration temperature | 21°C ±2°C (70°F ±4°F) |
| Sealing | IP54 (NEMA 4) |
| Physical size | 300 x 170 x 140 mm (11.8 in x 6.7 in x 5.5 in) |
| Weight | 3 kg (6.6 lb) |
Source: Druck DPI 610/615 Portable Pressure Calibrator Series User Manual
All Druck DPI 610/615 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What batteries does the DPI 610/615 use?
It uses 6 x 1.5 V C cells (alkaline), with up to 60 hours nominal use at 20°C, or a rechargeable NiCad battery pack with about 20 hours nominal use.
From: Druck DPI 610/615 Portable Pressure Calibrator Series User Manual
What is the operating temperature range?
The operating temperature range is -10°C to 50°C (+14°F to 122°F).
From: Druck DPI 610/615 Portable Pressure Calibrator Series User Manual
Can the DPI 615 load preset calibration and test procedures from a computer?
Yes, the DPI 615 can load preset calibration and test procedures from a computer.
From: Druck DPI 610/615 Series Portable Pressure Calibrator User's Manual
What is the size of the instrument?
The size is 300 x 170 x 140 мм.
From: Druck DPI 610/615 Series Portable Pressure Calibrator User's Manual
What pressure range does the DPI 610/615 cover?
The DPI 610/615 series can measure from 2.5 mbar up to 700 bar.
From: Druck DPI 610/615 Portable Pressure Calibrators User Manual
How many measurement values can be stored in the device?
The calibrator can store up to 10 000 measurement values in its internal memory.
From: Druck DPI 610/615 Portable Pressure Calibrators User Manual