User manual for the Druck DPI 610/615 portable pressure calibrator series, covering safety, specifications, operation, tasks, memory, setup, and calibration.
This user manual covers the Druck DPI 610/615 portable pressure calibrator series, including safety information, technical specifications, operator controls, display functions, electrical connections, and getting-started procedures.
It also describes task functions for pressure, voltage, and current measurements and outputs, advanced process functions, memory and datalog operations, setup options, calibration procedures, and hydraulic or low-pressure calibrator versions.
| Safe working pressure, 20 bar range | 1.75 x full-scale |
|---|---|
| Safe working pressure, 350 bar range | 1.2 x full-scale |
| Safe working pressure, 400 bar range | 1.5 x full-scale |
| Safe working pressure, all other ranges | 2 x full-scale |
| Accuracy | Combined non-linearity, hysteresis and repeatability: ±70 mbar range 0.05% F.S.; up to ±150 mbar 0.05% span; 200 mbar to 20 bar [Calibrator] 0.025% F.S.; 35 bar to 700 bar [Indicator] 0.025% F.S.; 70 bar to 400 bar [Hydraulic] 0.025% F.S. |
| Temperature effects | ±0.004% of reading/°C averaged over -10° to +40°C w.r.t. 20°C; ±0.002% of reading/°F averaged over +14° to 104°F w.r.t. 68°F |
| Power supply | 6 x 1.5 V C cells, alkaline (up to 60 hours nominal use at 20°C); rechargeable NiCad battery pack (20 hours nominal use) with charger/adaptor |
| Voltage inputs | Range ±50V; accuracy ±0.05% rdg, ±0.004% F.S.; resolution 100µV max |
| Voltage outputs | Range ±10V, accuracy ±0.1%, load 10mA; range ±24V, accuracy ±5%, load 26mA |
| Current inputs | Range ±55mA; accuracy ±0.05% rdg, ±0.004% F.S.; resolution 1µA max |
| Current output | Range 24mA; accuracy ±0.05% rdg, ±0.01% F.S.; resolution 1µA max |
| Display | 60 x 60 mm (2.36 in x 2.36 in) LCD graphics; reading ±99999; update rate 2 readings/sec |
| Operating temperature | -10°C to 50°C (+14°F to 122°F) |
| Calibrated temperature | -10°C to 40°C (+14°F to 104°F) |
| Storage temperature | -20°C to 60°C (-4°F to 140°F) |
| Calibration temperature | 21°C ±2°C (70°F ±4°F) |
| Sealing | IP54 (NEMA 4) |
| Physical size | 300 x 170 x 140 mm (11.8 in x 6.7 in x 5.5 in) |
| Weight | 3 kg (6.6 lb) |
It uses 6 x 1.5 V C cells (alkaline), with up to 60 hours nominal use at 20°C, or a rechargeable NiCad battery pack with about 20 hours nominal use.
The operating temperature range is -10°C to 50°C (+14°F to 122°F).
The instrument is sealed to IP54 (NEMA 4).
The display is a 60 x 60 mm LCD graphics display with a reading of ±99999 and an update rate of 2 readings/sec.