Aosong AO-09 Oxygen Sensor Product Manual

This manual provides technical specifications, installation instructions, and usage precautions for the Aosong AO-09 medical oxygen sensor.


Aosong AO-09 Oxygen Sensor Product Manual - cover page
Brand
AOSONG
Document type
User Manual
Language
English
Pages
6
File format
PDF
File size
281 KB
Published
26 October, 2026
Updated
26 October, 2026
MD5 checksum
F3B4769E961384EE630B9000BCE4C6E3

About this manual

The Aosong AO-09 is a high-quality replacement oxygen sensor designed specifically for medical applications. It features a molded body design with linear output from 0% to 100%, temperature compensation, and excellent signal stability without requiring an external power supply. The sensor meets ISO 80601-2-55 standards and is intended for use in oxygen analyzers, ventilators, anesthesia equipment, and incubators.

What's inside

  • Features
  • Applications
  • Technical Specifications
  • Product Dimensions
  • Installation Requirements
  • Stabilisation Time
  • Cleaning
  • Calibration Interval

Specifications

Operating PrinciplePartial pressure electrochemical
Output Range9 mV to 13 mV in 210 mBar O2
Measurement Range0 - 100 Vol.% O2
Response Time (T90)<15 s
Accuracy Standardmeets ISO 80601-2-55 requirements
Connector Type3P2.5
Housing MaterialWhite ABS
Weight36 g (nominal)
Operating Temperature Range0℃ ~ +50℃
Expected Operating Life (20°C)1.5 x 10^6%O2 hours

Frequently asked questions

What are the intended applications for the AO-09 sensor?

The sensors are intended as oxygen-sensing components for medical oxygen analyzers, control devices of oxygen concentrators, medical ventilators, anaesthesia equipment, and incubators.

Can the AO-09 sensor be sterilized by steam?

No, the sensor is not suitable for sterilization by steam or exposure to chemicals such as ethylene oxide or hydrogen peroxide.

What should be done if the sensor is dropped?

If a sensor is dropped, it should be placed in quarantine for 24 hours and followed up with a 2-point calibration.

Is the AO-09 suitable for MRI environments?

No, the sensor is not suited for use in a magnetic resonance imaging (MRI) environment.