Aosong AO-08 Oxygen Sensor User Manual
This manual provides instructions for installing, using, and maintaining the AO-08 oxygen sensor, designed for medical equipment such as ventilators and anesthesia machines. It details technical specifications, calibration procedures, and safety precautions, including warnings against misuse or exposure to harmful substances. The sensor features a linear output, temperature compensation, and a 13-month warranty.
The manual also includes mechanical and environmental specifications, cross-sensitivity test data, and guidelines for cleaning and handling. Users are advised to follow manufacturer instructions for calibration and to avoid steam sterilization or chemical exposure. The sensor is not suitable for MRI environments or applications involving personal injury risk. This manual is written in Chinese.
Aosong AO-08 specifications
| Operating principle | Partial pressure electrochemical |
|---|---|
| Output voltage | 9~13 mV |
| Measurement range | 0~ 100% |
| Response time (T90) | <15s |
| Accuracy | +1% |
| Linearity | Linear 0% to 100% O2 |
| Operating pressure range | 0.5 ~ 2.0 Bar |
| Expected operating life (20°C) | 1.5 x 10^6 % O2 hours |
| Shelf life | 13 months from date of dispatch |
All Aosong AO-08 documents by type
Ratings, dimensions and ordering codes, for selection and design work.
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the operating principle of the AO-08 sensor?
The sensor operates on the partial pressure electrochemical principle.
Does the AO-08 sensor require an external power supply?
No, the sensor requires no external power supply.
What is the maximum operating temperature for the AO-08 sensor?
The sensor is designed to operate within a temperature range of 0℃ to +50℃.
How long must the sensor stabilize in the instrument before calibration?
You must allow at least 15 minutes for the sensor to stabilize before calibrating.