Chinese user manual for the YTC YTC3000 earth resistance tester covering 4/3/2-wire ground resistance, soil resistivity, earth and AC voltage measurement.
This is the Chinese-language user manual for the YTC YTC3000 earth resistance tester, an instrument designed for field measurement of earth resistance, soil resistivity, earth voltage and AC voltage. It uses digital and microprocessor technology and supports precise 4-wire and 3-wire methods as well as a simplified 2-wire method, with FFT and AFC (automatic frequency control) technology to identify interference and select the measurement frequency.
The manual covers safety requirements, warranty terms, technical specifications, instrument structure and LCD indicators, measurement principles, and operating procedures including power on/off, battery voltage checks and AC voltage testing. The instrument stores 300 sets of data and can upload them to a computer via USB with monitoring software for history queries, reports and printing. This manual is written in Chinese.
| Earth resistance range | 0.00Ω~30.00kΩ |
|---|---|
| Soil resistivity range | 0.00Ωm~9000kΩm |
| Earth voltage range | 0V~600V |
| Power supply | DC9V (6× LR14 1.5V alkaline batteries, about 300 hours standby) |
| Test frequency | 128Hz/111Hz/105Hz/94Hz (AFC) |
| Short-circuit test current | AC 20mA max |
| Open-circuit test voltage | AC 40V max |
| Display | 4-digit large LCD, 128mm×75mm, white backlight |
| Data storage | 300 sets |
| Communication interface | USB, 1.5m cable, software monitoring |
| Dimensions | 212mm×175mm×76mm |
| Weight | Meter 950g; test leads 1560g; auxiliary ground rods 935g (4 rods) |
| Operating temperature/humidity | -10℃~40℃; below 80%rh |
| Insulation resistance | 20MΩ or more (500V between circuit and case) |
| Safety standards | IEC61010-1 (CAT III 300V, CAT IV 150V), IEC61557-1/-5, IEC61326 (EMC) |
It supports precise 4-wire and 3-wire measurement and a simplified 2-wire measurement of earth resistance.
It can store 300 sets of data; the display shows "FULL" when storage is full.
It is powered by DC 9V from six LR14 1.5V alkaline batteries, giving about 300 hours of continuous standby.
Soil resistivity is measured with the 4-pole (Wenner) method, calculated as ρ=2πaR with electrode spacing a between 1m and 100m, and the spacing should satisfy a>20h.