This manual provides operating instructions, safety guidelines, and technical specifications for the YTC YTC4206 Insulating Oil Gas Content Tester.
The document is a user operation manual for the YTC4206 Insulating Oil Gas Content Tester. It details the instrument's features, which include a 32-bit ARM MCU processor and compliance with DL/T423-2009 standards for measuring total gas content in insulating oil. This manual is written in Chinese.
| Working Power Supply | AC220V±10% 50Hz±1Hz |
|---|---|
| Maximum Power Consumption | 300W |
| Measurement Range | 0.2%~15% |
| Sample Volume | 100mL (1~3 times) |
| Degassing Chamber Volume | 250mL |
| Operating Temperature Range | 0~40℃ |
The instrument is developed based on the A method of DL/T423-2009 "Insulating Oil Gas Content Determination Method - Vacuum Pressure Difference Method".
The instrument calculates gas content by measuring the pressure difference generated when dissolved gas is released from a specific volume of oil under vacuum.
The instrument has external dimensions of 30×50×38cm and a host weight of approximately 15Kg.