User manual for the X-Rite i1Pro spectrophotometer, covering setup, calibration, measurement modes, maintenance, troubleshooting, specifications, and system…
The X-Rite i1Pro User Manual covers the i1Pro spectrophotometer, including package contents, device overview, installation, status indicators, calibration, and measurement procedures such as scan, spot, display, ambient light, and projector measurements.
It also describes regular maintenance, troubleshooting, warranty and support information, technical specifications, and minimum system requirements for Macintosh and Windows. This manual is written in Korean.
| Spectral engine | i1® technology (holographic diffraction grating with 128 pixel diode array) with built-in wavelengths check |
|---|---|
| Spectral range | 380 - 730 nm |
| Physical sampling interval | 3.5 nm |
| Optical resolution | 10 nm |
| Spectral reporting | 380 nm ... 730 nm in 10 nm steps |
| Measurement frequency in scanning mode | 200 measurements per second |
| Measurement geometry | 45º/0º ring illumination optics, ISO 13655:2009 |
| Measurement aperture | 4.5 mm (0.18”) diameter |
| Illumination spot size | 3.5 mm (0.14”) |
| Light source | Gas filled Tungsten (illuminant type A) and UV LED |
| Reflectance measurement | spectral reflectance [dimensionless] |
| Measurement conditions | UV included - ISO 13655:2009 measurement condition M0 D50 - ISO 13655:2009 measurement condition M1 UV excluded Filter - ISO 13655:2009 measurement condition M2 |
| OBC | Optical Brightener Compensation (OBC) with i1Profiler software |
| Calibration | Manual on external ceramic white reference |
| Measurement background | white, ISO 13655:2009; for measurements on backup board |
| Maximal media thickness | 3 mm (0.12”) on backup board |
| Minimal patch size in scanning mode | 7 x 10 mm (0.28” x 0.39”) (Width x Height) with sensor ruler; 10 x 10 mm (0.39” x 0.39”) (Width x Height) without sensor ruler |
| Inter-Instrument-Agreement | 0.4 ∆E94* average, 1.0 ∆E94* max. |
| Short-term repeatability | 0.1 ∆E94* on white (D50,2°, mean of 10 measurements every 3 s on white) |
| Emission measurement | spectral radiance [mW/nm/m 2/sr], luminance [cd/m2] |
| Measurement range | 0.2 - 1200 cd/m2 on a typical LCD-Monitor |
| Short-term repeatability | x,y ±0.002 typ. (5000 K, 80 cd/m2) |
| Ambient light measurement | spectral irradiance [mW/nm/m 2], illuminance [lux] |
| Interface | USB 1.1 |
The USB interface cable (2.0 m) shipped with this device must be used to maintain compliance.
The spectral range is 380 - 730 nm.
The measurement geometry is 45º/0º ring illumination optics, ISO 13655:2009.
Windows XP, Windows Vista, or Windows 7 with an Intel Pentium 4 or AMD Athlon XP or better CPU.